Pressure Relief 3 -
Therapeutic Surface Comparison
St. Mary's Hospital
Rochester, Minnesota

On June 8, 2005, Comfortex assisted with an Interface Pressure Evaluation with staff members from St. Mary’s Hospital (part of the Mayo system). The volunteer, a healthy
female, was used to compare the pressure reducing capabilities of a zoned, air adjustable bed and Comfortex’s non-mechanical Rest-Q mattress.

Interface pressures were recorded on a computer assisted X-SENSOR pressure monitor. The tests images were in a supine flat and Head Of Bed elevated position at 30 degrees & 45 degrees. A pillow was positioned beneath the head when in the elevated position.

Pressure relief chart
  Heel Area Sacral Area Shoulder Area
Zone Air System (flat)      
Readings over 33 mmHg 17 33 6
Readings over 52 mmHg 12 0 0
HOB@30 degrees      
Readings over 33 mmHg 19 124 28
Readings over 52 mmHg 14 17 4
HOB@45 degrees      
Readings over 33 mmHg 19 125 41
Readings over 52 mmHg 14 52 4
Rest-Q mattress (flat)      
Readings over 33 mmHg 14 32 0
Readings over 52 mmHg 6 0 0
HOB@30 degrees      
Over 33 mmHg 8 73 0
Over 52 mmHg 3 1 0
HOB@45 degrees      
Over 33 mmHg 9 158 0
Over 52 mmHg 3 3 0
 